Sequencer procedure: Difference between revisions

From SweepMe! Wiki
Jump to navigation Jump to search
m (fix link syntax)
(Added Sequencer procedure Navbox)
Line 98: Line 98:
* When a module gets active, i.e. it is part of an active [[branch]], it runs [[configure]] and [[poweron]]. If a module is not active anymore it runs [[poweroff]] and [[unconfigure]].
* When a module gets active, i.e. it is part of an active [[branch]], it runs [[configure]] and [[poweron]]. If a module is not active anymore it runs [[poweroff]] and [[unconfigure]].
* At each measurement point, i.e. a combination of setpoint iteration steps of all modules in a branch, a couple of functions are called to perform the measurement point such as [[start]], [[apply]], [[reach]], [[adapt]], [[measure]], [[call]] to name the most important ones.
* At each measurement point, i.e. a combination of setpoint iteration steps of all modules in a branch, a couple of functions are called to perform the measurement point such as [[start]], [[apply]], [[reach]], [[adapt]], [[measure]], [[call]] to name the most important ones.
{{Sequencer_procedure}}

Revision as of 11:46, 17 April 2024

This section provides an easy-to-understand overview of the functions executed in a SweepMe! Module. These functions are forwarded to a Driver if the Module can load drivers.

SweepMe! version 1.5.4 or higher

Sequencer Start:
connect() For all items from root to leaf: get and open a port object to communicate with a device.
initialize() For all items from root to leaf: Set values and configurations for devices to prepare for measurement.
---------------------------------------------------------------------------------------------------
If a new branch is entered and the module was not part of the previous branch:
configure() For all items from root to leaf: Set device configuration.
poweron() For all items from root to leaf: Powers on the device.
If an item of the sequencer starts its parameter variation:
signin() Calls reconfigure() if needed. Can be used to perform actions before parameter variation starts.
-------------------------------------------------------------------------------------------------
For each measurement point:
Each step is performed for all items from root to leaf before the next step is started:
start() Preparation before applying a new value.
----------------------------------------------------------------------------------------------
apply() Only called if the sweep value changes: Sets a new value to a device.
reach() Only called if apply() was called: Wait to reach the new value or start reaching a certain condition
sleephold() Sleeps during hold time while set values are applied.
----------------------------------------------------------------------------------------------
adapt() Adapts measurement devices to new conditions.
adapt_ready() Ensures all devices have adapted.
trigger_ready() Prepares devices to be triggered.
trigger() Deprecated, use measure() instead.
----------------------------------------------------------------------------------------------
measure() Initiates/Triggers a measurement.
request_result() Requests a result from the device buffer.
read_result() Reads a result from the port buffer.
----------------------------------------------------------------------------------------------
process_data() Processes measured data after the measurement finishes.
call() Mandatory: Returns values to SweepMe! as defined by self.variables.
----------------------------------------------------------------------------------------------
process() Only for Modules: Perform postprocessing after measurement data was obtained from the driver. Drivers should use process_data() to process data before it is transferred to SweepMe!.
finish() Cleans up before the next measurement.
--------------------------------------------------------------------------------------------------
If an item of the sequencer ends its parameter variation:
signout() Optionally performs actions after parameter variation.
Leaving a branch and the module is not part of the next branch or run is stopped by error:
poweroff() For all items from root to leaf: Switch off modules and devices.
unconfigure() For all items from root to leaf: Unconfigure modules and devices.
------------------------------------------------------------------------------------------------------
End of sequencer or if run is stopped by error:
deinitialize() For all items from root to leaf: Resets the status for any other user of the equipment.
disconnect() For all items from root to leaf: Closes the port object.



General rules

  • At the beginning of a measurement, modules and drivers use connect and initialize. At the end deinitialize and disconnect to bring the module or the driver back to an idle state and to close the connection.
  • When a module gets active, i.e. it is part of an active branch, it runs configure and poweron. If a module is not active anymore it runs poweroff and unconfigure.
  • At each measurement point, i.e. a combination of setpoint iteration steps of all modules in a branch, a couple of functions are called to perform the measurement point such as start, apply, reach, adapt, measure, call to name the most important ones.